Aquilion ONE / GENESIS Edition goes beyond the evolution of dynamic volume CT. Intensive, clinically focused research and innovative technological developments have culminated in a CT system with 16 cm z-axis coverage at 0.5 mm resolution and reduced radiation dose requirements¹.
Now available on Aquilion ONE / GENESIS Edition, Advanced intelligent Clear-IQ Engine (AiCE) is an innovative approach to CT reconstruction that leverages Deep Learning Neural Networks specifically trained to perform one task—reconstruct images that are sharp, clear, and distinct.
¹When AIDR 3D Enhanced and or FIRST Model-based Iterative Reconstruction are used.
A run off CTA was performed with AIDR3D demonstrating stents in the femoral arteries bilaterally.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 120 |
mAs | SUREExposure |
Rotation Time | 0.5 s |
Scan Range | 1457.0 mm |
Reconstruction | |
CTDlvol | 8.3 mGy |
DLP | 1287.0 mGy·cm |
A chest CT with iodinated contrast was performed with AiCE. It demonstrates lung and mediastinum masses.
View Scan ParametersScan Mode | Ultra-Helical |
Collimation | 0.5 mm x 80 |
kVp | 100 |
mAs | SUREExposure |
Rotation Time | 0.275 s |
Scan Range | 363.0 mm |
Reconstruction | |
CTDlvol | 3.8 mGy |
DLP | 156.0 mGy·cm |
Effective Dose* | 2.1 mSv |
*AAPM Report 96, k-factor 0.014
Courtesy of Prof. Prokop, Radboud UMC, Nijmegen, the Netherlands
Multiple tumors and evidence of COPD.
*Optional
View Scan ParametersNon-Contrast | Post-Contrast | |
Scan Mode | Ultra-Helical | Ultra-Helical |
Collimation | 0.5 mm x 80 | 0.5 mm x 80 |
kVp | 100 | 100 |
mAs | SUREExposure | SUREExposure |
Rotation Time | 0.275 s | 0.275 s |
Scan Range | 295.0 mm | 295.0 mm |
Reconstruction | ||
CTDIvol | 0.8 mGy | 2.7 mGy |
DLP | 30.9 mGy.cm | 97.7 mGy.cm |
Effective Dose* | 0.4 mSv | 1.3 mSv |
*AAPM report 96; k Factor 0.014